Product Details
  • Product Name:UV energy meter

  • Product number:UVICURE Plus Ⅱ
  • Product manufacturer:USA EIT
  • Product documentation:
你添加了1件商品 查看购物车
a brief introdction:
Single-channel UV meter Energy meter is easy to use, one click to complete ON/OFF and running operations, easy to read the data display, display 4 bands at the same time. Data from the four bands on the UV Power Puck II can be displayed simultaneously on the display for quick reading by the operator. The single-channel UV meter Energy meter, which first set the standard in the UV industry, is now
Details description:

Product picture


Product parameter

Single channel UV meter standard EIT multi-band one:
UVA (320-390nm) (default UVA), UVB (280-320nm)
UVC (250-260nm), UVV (395-445nm)
Dynamic range
Standard version - 10 Watt UVA, UVB, UVV; 1 Watt UVC. Low power version - 100 mW.
Setting function
Provides user selectable meter modes for data analysis, comparison, filtering and operation Settings.
Data reference mode
Used to compare readings. It is very useful in system installation and troubleshooting. The user can store the selected UV reading as a baseline or reference reading and then compare it to another reading. The meter displays two readings and indicates the percentage change between the readings. The data is shown as mJ/cm2,mW/cm2, and percentage.
The graphic mode displays the captured UV illuminance and energy for each UV band. The graph shows illuminance changing over time. The figure shown on the right represents a curing system with one or two lamps.
User selectable sampling rate
Smooth On Data: And the Power Puck before? Version compatible.
Smooth Off Data: Compatible with previous UV PowerMap over 2000 samples/second.
The units of measurement are user selectable, making it easy for the operator to read. The data will be displayed according to your needs. The selected units can be: mJ/cm2, mW/cm2, J/cm2, W/cm2, uJ/cm2, uW/cm2.
Colorful, easy to read display
Low, medium and high intensity can be selected for graphic display.
Communication port
UV energy meter and PC/PDA comply with serial communication protocol. Download the collected data to the computer for statistical analysis and data recording, as well as process verification.
Technical specification sheet
Display: Easy to read, yellow digits, black background
Measuring range: Standard version: UVA, UVB, UVV-10mW /cm2 to 10W/ m2; UVC -5mW/cm2 to 1W/cm2
Low power versions: UVA, UVB, UVC, UVV: 100microW/cm2 to 100mW/cm2
Measurement accuracy: +/ -10%; +/- 5% typical
Spectral range: (UVICURE Plus II)
1 channel continuous monitoring: 320-390nm (UVA), 280-320nm (UVB), 250-260nm(UVC), 395-445nm (UVV) (UV band must be determined before ordering)

Warm reminder

Spatial response: approximate cosine
Operating temperature: 0-75°C internal temperature. Allow higher external temperatures for a short period of time (an alarm will sound when the temperature exceeds the corresponding specification)
Time: 2 minutes DISPLAY mode (no key action). EITIM can also be set to display no time.
Batteries: Two replaceable AAA alkaline batteries
Battery life: About 20 hours display time
Dimensions: Diameter 117mm x height 127mm
Weight :289 grams
Housing material: aluminum and stainless steel
Handbag Material: Internal polyurethane, soft, scratch-resistant nylon shell
Weight: 260g
Handbag size: 274mm wide x 89mm high x 197mm deep